Chapter 2 Performance Tests
58502A Performance Test Record (Page 1 of 2)
2-14 User’s Guide
58502A Performance Test Record
(Page 1 of 2)
Model 58502A
Broadband Distribution Amplifier
Serial Number:_______________________________ Repair/Work Order No.__________________
Test Performed By:___________________________ Temperature:___________________________
Date:________________________________________ Relative Humidity:______________________
Notes: ______________________________________________________________________________
Line Number Operational Verification Test Results
Pass Fail
1 Outputs (1-12) Verification
____ ____
2 Alarm/Status Verification
____ ____
3 RS-232 Verification
____ ____
58502A Performance Tests
Input Level Response Tests
Test 1: +7 dBm Input Level
Minimum Actual Reading (Outputs 1 – 12) Maximum
0.707 Vrms < 1. ________________________ < 1.12 Vrms
0.707 Vrms < 2. ________________________ < 1.12 Vrms
0.707 Vrms < 3. ________________________ < 1.12 Vrms
0.707 Vrms < 4. ________________________ < 1.12 Vrms
0.707 Vrms < 5. ________________________ < 1.12 Vrms
0.707 Vrms < 6. ________________________ < 1.12 Vrms
0.707 Vrms < 7. ________________________ < 1.12 Vrms
0.707 Vrms < 8. ________________________ < 1.12 Vrms
0.707 Vrms < 9. ________________________ < 1.12 Vrms
0.707 Vrms < 10. _______________________ < 1.12 Vrms
0.707 Vrms < 11. _______________________ < 1.12 Vrms
0.707 Vrms < 12. _______________________ < 1.12 Vrms